Bio-SiC ceramics coated with hydroxyapatite using gas-detonation deposition: An alternative to titanium-based medical implants

dc.contributor.authorKlyui, M.I.
dc.contributor.authorTemchenko, V.P.
dc.contributor.authorGryshkov, O.P.
dc.contributor.authorDubok, V.A.
dc.contributor.authorKladko, V.P.
dc.contributor.authorKuchuk, A.V.
dc.contributor.authorDzhagan, V.M.
dc.contributor.authorYukhymchuk, V.O.
dc.contributor.authorKiselov, V.S.
dc.date.accessioned2017-06-10T13:24:06Z
dc.date.available2017-06-10T13:24:06Z
dc.date.issued2013
dc.description.abstractSilicon-carbide (SiC) ceramics synthesized using forced infiltration and coated with bioactive hydroxyapatite (HA), represents a great potential for replacement of traditional titanium medical implants. In this paper the surface morphology, structural properties and phase content of HA coatings deposited onto nano-porous bio-SiC ceramics with pore orientation being parallel or perpendicular to the plain of subsequently deposited HA film have been studied. The HA coatings produced at 5 s of deposition contained cracks, while the uniform hydroxyapatite coating has been produced at 10 s of deposition. X-ray diffraction analysis revealed that the HA coatings preserve crystal structure of the initial hydroxyapatite powder during deposition. From the intensity ratio of Raman scattering peaks, corresponding to crystalline (967 cm⁻¹) and amorphous (951 cm⁻¹) HA, the least and most crystalline structures were found. The structural perfection of the HA coating was higher for SiC substrates with parallel pore orientation (with respect to the plane of HA growth), indicating that structural properties of SiC ceramics affect the morphological and structural perfection of deposited HA coating.uk_UA
dc.identifier.citationBio-SiC ceramics coated with hydroxyapatite using gas-detonation deposition: An alternative to titanium-based medical implants / M.I. Klyui, V.P. Temchenko, O.P. Gryshkov, V.A. Dubok, V.P. Kladko, A.V. Kuchuk, V.M. Dzhagan, V.O. Yukhymchuk, V.S. Kiselov // Functional Materials. — 2013. — Т. 20, № 2. — С. 163-171. — Бібліогр.: 24 назв. — англ.uk_UA
dc.identifier.issn1027-5495
dc.identifier.otherDOI: dx.doi.org/10.15407/fm20.02.163
dc.identifier.urihttps://nasplib.isofts.kiev.ua/handle/123456789/119954
dc.language.isoenuk_UA
dc.publisherНТК «Інститут монокристалів» НАН Україниuk_UA
dc.relation.ispartofFunctional Materials
dc.statuspublished earlieruk_UA
dc.subjectCharacterization and propertiesuk_UA
dc.titleBio-SiC ceramics coated with hydroxyapatite using gas-detonation deposition: An alternative to titanium-based medical implantsuk_UA
dc.typeArticleuk_UA

Файли

Оригінальний контейнер

Зараз показуємо 1 - 1 з 1
Завантаження...
Ескіз
Назва:
06-Klyui.pdf
Розмір:
1.75 MB
Формат:
Adobe Portable Document Format
Опис:
Стаття

Контейнер ліцензії

Зараз показуємо 1 - 1 з 1
Завантаження...
Ескіз
Назва:
license.txt
Розмір:
817 B
Формат:
Item-specific license agreed upon to submission
Опис: